Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add filters








Year range
1.
Rev. estomatol. Hered ; 31(1): 37-43, ene-mar 2021. tab, graf
Article in Portuguese | LILACS-Express | LILACS | ID: biblio-1251765

ABSTRACT

RESUMEN Objetivo: Realizar, durante siete años ininterrumpidos, un control y seguimiento de las fuentes de luz disponibles en las clínicas odontológicas de pre graado de una Institución de Enseñanza Superior Pública Brasileña. Material y métodos: Durante el período comprendido entre 2011-2017 se evaluó el control de calidad de las fuentes de luz de la Facultad de Odontología de la Universidad Federal de Goiás a través del análisis de su estado de conservación y de su densidad de potencia (mW/cm2). Los datos obtenidos se sometieron a un análisis estadístico descritpivo y a las pruebas de Levene, Shapiro-Wilk, Kruskal-Wallis, Dunn y Chi-cuadrado (P<0,05). Resultados: No se observaron diferencias estadísticas significativas del estado de conservación de las fuentes de luz entre los años 2012, 2016 y 2017 (p>0,05). Sí se observaron diferencias estadísticas significativas entre los valores de densidad de potência entre dichos años (p<0,0001). En las comparaciones por pares, los valores de densidad de potencia para los años 2011 (p<0,01) y 2012 (p<0,05) fueron estadísticamente diferentes de los años 2015, 2016 y 2017. Conclusiones: Comparativamente se observó una mejora sustancial del estado de conservación de las fuentes de luz en los cuatro años iniciales de evaluaciones, con una disminución en el quinto año y una estabilización en los dos últimos años. Durante los siete años de estudio de la densidad de potencia de las fuentes de luz se observó en un aumento gradual de la misma con una estabilización de los resultados a partir del quinto año de control de calidad.


SUMMARY Objetive: To monitor the light curing units available at undergraduate dental clinics of a Brazilian Public Higher Education Institution for seven uninterrupted years. Materials and methods: In the period between 2011-2017, the quality control of light curing units at the Faculty of Dentistry, Federal University of Goiás, was evaluated by analyzing their state of conservation and power density (mW/cm2). Data obtained were subjected to descriptive statistical analysis and Levene, Shapiro-Wilk, Kruskal-Wallis, Dunn and Chi-square tests (P <0.05). Results: Regarding the conservation status of the light curing units, comparatively, there was no statistically significant difference between the years 2012, 2016 and 2017 (p> 0.05). Statistically significant differences were found for power density values between years (p <0.0001). In pairwise comparisons, the power density values for the years 2011 (p <0.01) and 2012 (p <0.05) were statistically different from the years 2015, 2016 and 2017. Conclusions: A significant improvement in the state of conservation of light curing units was observed in the initial four years of evaluations, with a decline in the fifth year and stabilization in the last two years. Regarding the power density of the light curing units, a gradual increase was observed in the seven years of study, with stabilization of the results after the fifth year of quality control.

2.
Rev. patol. trop ; 48(2): 87-98, 2019.
Article in English | LILACS | ID: biblio-1025775

ABSTRACT

The work conditions of dental surgeons (DS), associated with low compliance to precautionary standards, often lead to the colonization and dissemination of infectious agents. To assess the epidemiological and microbiological aspects of nasal colonization by Gram-negative bacteria in DS while teaching (dentistry teachers). The data were collected by application of a questionnaire and a nasal swab. The biochemical identification, the susceptibility profile and the detection of ß-lactamases were carried out in Vitek 2 compact®. 41 (77.3%) DS participated in the study, nine of them (22.0%) presented nasal colonization by Enterobacteriaceae, the participants were predominantly men (27/65.9%), over 50 years of age (24/58.5%). All of them confirmed using procedure gloves and a surgical mask while attending patients, with frequent (95.5%) hand washing although a statistical difference was found regarding personal habits (p=0.03). Enterobacter aerogenes (60.0%) was the most prevalent species, followed by Citrobacter koseri (20.0%). The intrinsic production of AmpC ß-lactamase by E. aerogenesspecies, which is multiresistant to antimicrobials, was present in the nasal cavity of 14.6% of the DS. There were high levels of nasal colonization by Enterobacteriaceae in teaching DS (22.0%), 14.6% had been colonized by multiresistant microorganisms and the results were associated with inadequate personal habits.


Subject(s)
Humans , Occupational Risks , Dentistry , Gram-Negative Bacteria , Nasal Cavity
SELECTION OF CITATIONS
SEARCH DETAIL